Temperature limits, 1 field versions – KROHNE OPTIFLUX 2000-4000-5000-6000 -IFC 100 Ex EN User Manual
Page 14
2
TEMPERATURE LIMITS
14
OPTIFLUX
www.krohne.com
04/2009 - 7.30685.22.00 - EX IFC 100 en
2.2 OPTIFLUX 5000
The maximum process temperature T
p
is dependent on the required temperature class T6...T3
and the maximum ambient temperature T
a
.
2.2.1 Field versions
Sandwich versions
• The flow sensor is suitable for an ambient temperature range of -20...65°C (DN2.5...15) or -
40...+65°C (DN25...100).
• The minimum process temperature is -20°C (DN2.5...15) or -40°C (DN25...100).
Flange versions
• The flow sensor is suitable for an ambient temperature range of -40...65°C (DN15...100).
• The minimum process temperature is -40°C.
Temperature class
(for gasses)
Maximum surface
temperature for
dust [°C]
Max. process temperature Tp [°C]
T
a
≤ 40°C
40 < T
a
≤ 50°C
50 < T
a
≤ 65°C
T6
85
65
65
60
T5
95
85
85
75
T4
130
125
125
115
T3
180
180
165
140
Use heat
resistance cable
for T
p
above:
1
165
130
100
1 Cables must withstand a continuous operating temperature of 85 °C.
Temperature class
(for gasses)
Maximum surface
temperature for
dust [°C]
Max. process temperature Tp [°C]
T
a
≤ 40°C
40 < T
a
≤ 50°C
50 < T
a
≤ 65°C
T6
85
80
80
75
T5
95
95
95
95
T4
130
130
130
130
T3
180
180
180
145
Use heat
resistance cable
for T
p
above:
1
never
165
90
1 Cables must withstand a continuous operating temperature of 85 °C.