beautypg.com

Installation, 8 installation conditions, 1 inlet and outlet – KROHNE OPTIFLUX 1000 Quickstart EN User Manual

Page 8

background image

2

INSTALLATION

8

OPTIFLUX 1000

www.krohne.com

04/2014 - 4002298502 - QS OPTIFLUX 1000 R06 en

2.8 Installation conditions

2.8.1 Inlet and outlet

Use straight inlet and outlet pipe sections to prevent flow distortion or swirl, caused by bends
and T- sections

2.8.2 Bends in 2 or 3 dimensions

2.8.3 T-section

Figure 2-5: Recommended inlet and outlet section

1 Refer to chapter "Bends in 2 or 3 dimensions"
2 ≥ 2 DN

Figure 2-6: 2 and 3 dimensional bends, in front of flowmeter

1 Bends in 2 dimensions: ≥ 5 DN; bends in 3 dimensions: ≥ 10 DN

Figure 2-7: Distance behind a T-section

1 ≥ 10 DN