beautypg.com

Electrical connections – KROHNE IFC 100 Converter Quickstart EN User Manual

Page 14

background image

3

ELECTRICAL CONNECTIONS

14

IFC 100

www.krohne.com

07/2010 - 4000124104 - QS IFC 100 R04 en

3.4.3 Preparing signal cable A, connection to signal converter

• Connection of the two shields in the signal converter is carried out via the stranded drain

wires.

• Bending radius: ≥ 50 mm / 2"

Required materials

• PVC insulation tubing, Ø2.5 mm / 0.1"
• Heat-shrinkable tubing
• 2x wire end ferrules to DIN 46 228: E 1.5-8 for the stranded drain wires (1, 6)
• 2x wire end ferrules to DIN 46 228: E 0.5-8 for the insulated conductors (2, 3)

1 Strip the conductor to dimension a.
2 Cut off the inner shield (10) and the outer shield (60). Make sure not to damage the stranded

drain wires (1, 6).

3 Slide the insulation tubing over the stranded drain wires (1, 6)
4 Crimp the wire end ferrules onto the stranded drain wire.
5 Crimp the wire end ferrules onto the conductors (2, 3).
6 Pull the heat-shrinkable tubing over the prepared signal cable.

INFORMATION!
Assembly materials and tools are not part of the delivery. Use the assembly materials and tools
in compliance with the applicable occupational health and safety directives.

Figure 3-3: Preparation of signal cable A

a = 80 mm / 3.15"