Introduction
1.1 Instrument Description
The Model 3510 is a general purpose pH/mV/Temperature bench meter used for routine laboratory
analysis. The meter supports 1, 2 or 3 point pH calibration on either manually entered pH buffer
values or automatically temperature compensated buffers to DIN, JIS and NIST standards and
Jenway buffers supplied with the instrument. Up to 3 decimal place resolution is available. The 3510
includes a 32 reading memory facility.
1.2 Instrument Specification
pH (1, 2 or 3 point cal)
Range:
-2.000 to 19.999pH
Resolution:
0.001 / 0.01 / 0.1pH
Accuracy:
±0.003pH
mV (Absolute or Relative)
Range:
-1999 to +1999mV
Resolution:
0.1mV
Accuracy:
±0.2mV
Input Impedance:
>10
12
ohms
Temperature Measuring
Ranges:
-10 to +105°C / 14 to 221°F
Resolution:
0.1°C / 1°F
Accuracy:
±0.5°C / ±1°F
ATC Range:
0 to 100°C / 32 to 212°F
Manual Temp. Compensation:
0 to 100°C / 32 to 212°F
Auto Buffer Selection:
Jenway (2.00, 4.00, 7.00, 9.20 and 10.00)
DIN (3.06, 4.65, 6.79, 9.23, 12.75)
NIST (1.68, 4.01, 6.87, 9.18, 12.45)
JIS (1.68, 4.01, 6.87, 9.18, 12.45) or manually entered buffers
Calibration:
User selectable 1, 2 or 3 point
Outputs:
Analogue 1mV per 0.01pH
RS232
Display:
Back lit custom LCD
Power:
Power Supply 9Vac
Size:
275(l)x240(w)x150(d)mm
Weight:
850g
351 050/REV B/09-06