Jenway 3510 User Manual
Page 30
Section 7
Troubleshootiing and functional checks
7.1 Troubleshooting
Fault
Possible Cause
Action
No display
Check power supply
Check that correct 9V ac power supply
is connected and switched on.
Erratic display
Check power supply
Unit must be used with supplied 9V
acpower supply. Usage of other units will
cause the 3510 not to operate.
Drifting erratic readings
Electrode fault
Use BNC cap to test 3510 (see 7.2)
Replace electrode.
Cannot calibrate
Electrode Fault
Use BNC cap to test 3510 (see 7.2)
Replace electrode.
ERROR EO ...
Buffer problem
Use freshly prepared buffers.
ERROR SLOPE ....
Electrode problem
Use BNC cap to test 3510 (see 7.2)
Replace electrode.
Unable to recognise
Using correct buffer set
Is the buffer type correct? Use AUTO for
buffers
Jenway supplied buffers.
Replace buffers
Use BNC cap to test 3510 (see 7.2)
Replace electrode.
Temperature readings
Temp probe faulty
Check 3510 using section 7.2
fluctuating
Replace temperature probe.
Temperature readings
Temp probe faulty
Check 3510 using section 7.2
incorrect
Replace temperature probe.
Manual temp not set
Set meter to read °C and set
temperature against a calibrated
thermometer.
Will not print
Connection
3510 requires the RS232 cable to
connect the printer.
Paper out
The feed light on the printer will flash if
the unit requires paper.
Battery flat
Connect ac power supply.
If the above does not answer your query try the FAQ section on the
www.Jenway.com
Website.
351 050/REV B/09-06
27