INOR IPAQ-4L GB User Manual
Universal high-isolation 4-wire transmitter

MEASURE OF SUCCESS
USER INSTRUCTIONS
www.inor.com, www.inor.se
The manual must be read prior to adjustment
and/or installation. All information subject to
change without notice.
The transmitter is configured from a standard IBM compatible PC
by using the IPRO software (version 4.1 or later). IPRO 4.1 is com-
patible with Windows 98SE, Windows NT Workstation 4.0, Windows
2000 and Windows XP. When the transmitter is reconfigurated by a
PC no recalibration is needed.
Connect the IPRO communication cable between the serial port of
a PC (with IPRO software installed) and the PC connection port on
IPAQ-4L.
The IPAQ-4L must be powered, when you configure the
transmitter.
If IPRO (the IPAQ software) isn’t installed in your PC, install the
software (follow the instructions in ”IPRO”). Configuration from
PC is “on-line”, that is, the transmitter can be installed in the process
with input connected during configuration. The output is frozen while
transmission from PC to IPAQ take place. When transfer is done the
transmitter uses the new parameters.
NOTE: The label on the transmitter indicate allowed power supply
to the transmitter. By connecting false power supply the
transmitter may be damaged.
INTRODUCTION
IPAQ-4L is a fully universal and programmable 4-wire (separately
powered) transmitter. The high isolation provides safe and problem
free measurements. Offering both current and voltage output,
IPAQ-4L meets any process requirements. Designed for wall and
DIN-rail mounting.
IPAQ-4L is intended to be used in industrial environment.
DIMENSIONS
Universal
High-Isolation
4-wire Transmitter
CONFIGURATION
SWEDEN
GERMANY
USA
FINLAND
Since IPAQ-4L
is a precision instrument it´s recommended to check
the calibration once a year. Use equipment with an
inaccuracy of at least 5 times better than the checked one, e.g. if you
are checking a transmitter with an inaccuracy of 0.1% use instrument
with an inaccuracy of maximum 0.02%.
You can either perform a full recalibration of the transmitters stored
calibration parameters (Basic Calibration) or make a System
Error Correction (recommended). In each case use the IPRO software
and follow the instructions on the screen. When performing a System
Error Correction, please note that this is only a compensation for
errors at the measuring range in use.
CALIBRATION
86BIQ00007 2012-10
mm/inches
This pr
oduct should not be mix
ed wi
th other kind of scr
ap
, after usage. It should be handled as an electr
onic/electric device.
PC connection
ADDRESS 7 Dearborn Road,
Peabody, MA 01960
USA
PHONE
+1-978-826 6900
FAX
+1-978-535 3882
ADDRESS PO Box 9125
SE-200 39 Malmö
Sweden
PHONE
+46-40-31 25 60
FAX
+46-40-31 25 70
ADDRESS Am Bocksborn 5
D-63571 Gelnhausen
Germany
PHONE
+49-6051 14807
FAX
+49-6051 14806
ADDRESS Unikkotie 13
FI-01300 Vantaa
Finland
PHONE
+358-10-421 7900
FAX
+358-10-421 7901
INOR Process AB
INOR Transmitter GmbH
INOR Transmitter Oy
KROHNE Temperature
Division INOR