Warning – HT instruments XL424 User Manual

Page 4

background image

XL423 - XL424

EN - 3

1.2 DURING

USE

Carefully read the herewith recommendations and instructions:

WARNING

• Non compliance with warnings and/or instructions may cause damage to

the tester or its components or injure the operator.

• During a measurement never touch any unused terminal.



1.3 AFTER

USE

If you expect not to use the instrument for a long period of time remove the batteries

and strictly follow the indications at paragraph 9.2.


1.4 OVERVOLTAGE

CATEGORIES - DEFINITIONS

EN 61010-1 (Safety requirements for electrical equipment for measurement, control and
laboratory use, Part 1: General requirements) defines what a measurement category
(usually called “overvoltage category”) is. At paragraph 6.7.4: Measuring circuits it says:

(OMISSIS)

Circuits are divided into the following measurement categories:

Measurement category IV is for measurements performed at the source of low

voltage installations.

Examples are electricity meters and measurements on primary overcurrent protection devices
and ripple control units.

Measurement category III is for measurements performed in the building

installation.

Examples are measurements on distribution boards, circuit breakers, wiring,
including cables, bus-bars, junction boxes, switches, socket-outlets in the fixed
installation, and equipment for industrial use and some other equipment, for
example, stationary motors with permanent connection to fixed installation.

Measurement category II is for measurements performed on circuits directly

connected to the low voltage installation.

Examples are measurements on household appliances, portable tools and similar
equipment.

Measurement category I is for measurements performed on circuits not directly

connected to MAINS.

Examples are measurements on circuits not derived from MAINS, and specially
protected (internal) MAINS-derived circuits. In the latter case, transient stresses are
variable; for that reason, the norm requires that the transient withstand capability of
the equipment is made known to the user.


This manual is related to the following products: