HT9021
EN - 18
Capacitance (Autorange)
Range Resolution Accuracy
Overload
protection
40.00nF 0.01nF
±(3.5%rdg+40dgt)
600VDC/ACrms
400.0nF 0.1nF
±(2.5%rdg+5dgt)
4.000
µF 0.001µF
40.00
µF 0.01µF
400.0
µF 0.1µF
4000
µF 1µF
±(5.0%rdg+5dgt)
Diode test
Range
Test current
Open voltage
0.9mA typical
2.8VDC
Frequency with test leads (Autorange)
Range Resolution Accuracy Sensitivity
Overload
protection
99.99Hz 0.01Hz
±(1.0%rdg+5dgt)
≥15Vrms
600VDC/ACrms
999.9Hz 0.1Hz
9.999kHz 0.001kHz
60.00kHz 0.01kHz
Input current for frequency measurement with jaws: > 8% selectable full scale
Frequency with jaws (Autorange)
Range Resolution Accuracy Sensitivity
Overload
protection
99.99Hz 0.01Hz
±(1.0%rdg+5dgt)
≥10A (60A)
≥50A(600A)
1000VACrms
999.9Hz 0.1Hz
9.999kHz 0.001kHz
Duty Cycle (Autorange)
Range Resolution Sensitivity
Accuracy
0.5%
÷ 99.0%
0.1%
≥15Vrms
±(1.2%rdg+2dgt)
Pulse width: 100
µs ÷100ms ; Pulse frequency: 5.000Hz ÷100.0kHz
Temperature with type K probe (Autorange)
Range Resolution Accuracy
(*)
Overload
protection
-20.0
÷ 400.0°C
0.1°C
±(2.0%rdg +3°C)
600VDC/ACrms
400
÷ 760°C
1°C
±(2.0%rdg +5°C)
-4.0
÷ 752.0°F
0.1°F
±(2.0%rdg +6°F)
752
÷ 1400°F
1°F
±(2.0%rdg +9°F)
(*) Accuracy of type K probe not considered
6.1.1. Safety
Comply with:
IEC/EN 61010-1
Insulation: double
insulation
Pollution: Level
2
Max height of use:
2000m (6562 ft)
Installation category:
CAT IV 600V, CAT III 1000V to ground,
6.1.2. General data
Mechanical characteristics
Dimensions (L x W x H):
252 x 88 x 44mm ; 10 x 3 x 2 in
Weight (including battery):
442g (16 ounces)
Max conductor size:
45mm (2 in)