beautypg.com

HT instruments HT8100 User Manual

Page 23

background image

HT8100

EN - 22

Resistance

Range Resolution Accuracy

Output

current

Protection against

overcharge

500.00

 0.01

(0.2%rdg+30dgt)

1mA

1000VDC/ACrms

5.0000k

 0.0001k

(0.2%rdg+10dgt)

100µA

50.000k

 0.001k

10µA

500.00k

 0.01k

(0.5%rdg+10dgt)

1µA

5.0000M

 0.0001M

(1.0%rdg+10dgt)

100nA

50.00M

 (*)

0.01M

(2.0%rdg+10dgt)

10nA

(*) Slight instability < 20 dgt
Maximum open-circuit voltage: approx 3.5V

Continuity test

Range Accuracy

Buzzer

Open-circuit voltage

Protection against

overcharge

500.00

(0.1%rdg+30dgt) <30

circa 3.5V

1000VDC/ACrms

Diode test

Range Accuracy

Test

current

Open-circuit voltage

Protection against

overcharge

2.000V

(1.0%rdg+10dgt)

±1mA circa

±3V

1000VDC/ACrms

AC Voltage and AC Current Frequency

Range Resolution Accuracy

Protection against overcharge

500.00Hz 0.01Hz

3dgt

1000VDC/ACrms

max 440mA

5.0000kHz 0.0001kHz
50.000kHz 0.001kHz
100.00kHz 0.01kHz

Minimum frequency value: 5Hz


Sensitivity for frequency measurements

Function Range

Sensitivity (peak to peak value)

5Hz ÷ 10kHz

10kHz ÷ 100kHz

AC mV

50.000mV

10mV 100mV

500.00mV

AC V

5.0000V 1V

1V

50.000V

1V

Unspecified

500.00V
1000.0V

AC A

50.000mA 10mA

1.000A

300mA

Generated DC Current – Programmable output

Range Resolution Accuracy

Protection against

overcharge

0.000

20.000mA

0.001mA

(0.05%rdg+5dgt)

max 440mA

4.000

20.000mA

Power supply: battery level > 4.5V. External power supply in simulation mode: 6V ÷ 48V
Source mode: 1.200

 @20mA (battery voltage >4.5V)