Caution – HT instruments HT8100 User Manual
Page 14
HT8100
EN - 13
4.4.3. DC Current measurement
CAUTION
The maximum input DC current is 1A. Do not measure currents exceeding
the limits given in this manual. Exceeding current limits could result in
electrical shocks to the user and damage to the instrument.
Fig. 4: Use of the instrument for DC current measurement
1. Cut off power supply from the circuit to be measured.
2. Select position A
or mA
(for measuring currents <50mA).
3. Press the MODE key for manually selecting “DC” measurement.
4. Insert the red cable into input lead mA
or A
and the black cable into input lead
COM.
5. Connect the red lead and the black lead in series to the circuit whose current you want
to measure, respecting polarity and current direction (see Fig. 4)
6. Supply the circuit to be measured. The display shows the value of current.
7. If the display shows the message “OL”, the maximum measurable value has been
reached.
8. When symbol “-” appears on the instrument’s display, it means that current has the
opposite direction with respect to the connection in Fig. 4
9. For using the HOLD function see § 4.2.1, for measuring MAX/MIN/AVG values see
§ 4.3.1, for relative measurement see § 4.2.1 and for saving the result see § 4.3.5