1000-XCTV

High AC Current

Equipment Required:

1. Precision AC current source with a resolution of 0.1ma AC

and an accuracy of 0.1%.

2. Interconnecting copper wire from the AC voltage source to the

indicator.

3. Trimmer adjusting tool.

STEP 1.

Disconnect power from the unit and remove the front

panel lens. Carefully remove the electronics from the housing by
lifting the display above and over the latch located at the bottom
front of the housing and slide the assembly out of the case.

STEP 2.

Make note of the jumper settings on JP501. Carefully

remove them and replace jumpers on position 3 and position 10.

STEP 3.

Connect the AC current source to terminals 5 and 6

located on the rear of the unit. Reconnect power and turn the unit
on. Allow for a 5-minute warm-up.

STEP 4.

Input 0.0 amp from the AC current source and adjust

the ZERO pot for a display reading of 0 counts ±1 count.

STEP 5.

Input 1 amp @ 60 Hz from the AC current source and

adjust the SPAN pot for a display reading of 1000 counts ±1 count.

STEP 6.

Remove the jumper on positions 3 and 10 and replace

all of the jumpers to their original settings.