TO SET THE SCALE
Set the scale to
#
1 for most paper and paper products,
#
2 for
obtaining relative moisture indications on hygroscopic materials
for which no established calibration is available, or
#
3 for
baled scrap paper.
hTo change the scale, press and hold the scale key
$
. The
meter will display the current scale for one second then
scroll forward through the scales.
hRelease the key to stop at your desired scale.
Changing the scale will automatically reset the set-point value
to the default setting for that particular scale. Default settings
are as follows:
Scale
#
1 – 7%
Scale
#
2 – 50%
Scale
#
3 – 19%
TO CHANGE THE SET-POINT
hTo change the set-point value press the set-point key
#
.
The meter will display the current set-point value for the
scale you have chosen for one second.
hTo scroll forward to a higher value for that scale hold the
set-point key
#
while the current set-point is displayed
and scroll to the set-point value desired.
hTo scroll backward through the set-point values, press and
release the set-point key
#
. Within one second, press and
hold the scale key
$
. Continue to hold the scale key
$
and the set-point will decrease.
hWhen scrolling in either direction, release the key to stop
at your desired set-point.
hA buzzer sounds if the meter reads a value higher than the
set-point.
You can change the value between 5.0 and 18.0 for Scale
#
1,
2-99 for Scale
#
2, and between 6.0 and 39.0 for Scale
#
3.
4
Disconnect the battery. Press and hold the Read key for 15 sec-
onds. Release the Read key. Press and hold the Check key for 15
seconds. Release the Check key. Connect a fresh battery to the
lead wire in a single action, making sure to align the poles prop-
erly and without interrupting contact. If the display remains
frozen, repeat the procedure. If this procedure does not solve the
problem, refer to the Service for Your Meter section.