Delmhorst uses the USDA standard-Douglas Fir as the basis for all calibrations.
For any species other than Douglas Fir and for wood temperature out of the
range of 50°F (10°C) to 90°F (32°C) temperature and species adjustment must
be made. Correct for temperature first, then species. See temperature correction
table and species correction tables below.

The RC-1E can be used for more than just wood. It will also give a relative
reading on plywood, OSB, particleboard and MDF. Call Delmhorst at 877-
DELMHORST or e-mail

[email protected]

for information on how to interpret

the readings for other materials.


TEMPERATURE CORRECTION TABLE

METER READINGS

°C

°F

6

7

10

15

20

25

30

35

40

50

60

-20

0

9

11

15

22

31

38

45

53

-10

20

8

10

14

20

28

34

40

47

55

5

40

7

8

12

18

24

30

36

42

48

15

60

6

7

11

16

21

27

32

38

43

54

30

80

6

7

9

14

19

23

28

33

38

47

55

40

100

5

6

8

12

17

21

25

29

34

42

50

50

120

5

5

7

11

15

19

22

26

30

38

44

60

140

4

5

7

10

14

17

20

23

27

34

40

70

160

4

4

6

9

12

15

18

21

24

30

36

80

180

3

4

5

8

11

13

16

19

22

27

33

95

200

3

4

5

7

10

12

14

17

19

24

28

105

220

2

3

4

6

9

11

13

15

17

21

26


Moisture content values shown shaded

are only qualitative, since they are above

the fiber saturation point.

The temperature correction values shown in this chart have been rounded for
easy reference.