Delmhorst Instrument C-2000 User Manual
Page 6

A hard Reset is required if, after changing the battery, the
display is frozen. This is sometimes caused by the interrup-
tion of contact between the battery and the battery lead wire.
Resolve this as follows:
Disconnect the battery. Press and hold the Read key for 15
seconds. Press and hold the Check key for 15 seconds.
Release the Check key. Connect a fresh battery to the lead
wire in a single action, making sure to align the poles
properly and without interrupting contact. If the display
remains frozen, repeat the procedure. If this procedure
does not solve the problem, refer to the Service for Your
Meter section.
TO SET THE SCALE
Set the scale for the material being tested: #1 Lint cotton, #2
Seed cotton, #3 Viscose Rayon, #4 Wool.
h
To change the scale, press the scale key
$
. The meter will
display the current scale for one second.
h
Press and hold the scale key
$
to toggle through the four
scales.
h
Release the scale key
$
to stop at the desired scale.
Changing the scale will automatically reset the set-point value
to the default setting for that particular scale. Default settings
are as follows:
Lint cotton - 10%
Viscose Rayon - 18%
Seed cotton - 10%
Wool - 20%
TO CHANGE THE SET-POINT
h
To change the set-point value press the set-point key
#
.
The meter will display the current set-point value for the
scale you have chosen for one second.
h
To scroll forward to a higher value for that scale hold the
set-point key
#
while the current set-point is displayed
and scroll to the set-point value desired.
h
To scroll backward through the set-point values, press and
release the set-point key
#
. Within one second, press and
hold the scale key
$
. Continue to hold the scale key
$
and the set-point will decrease.
h
When scrolling in either direction, release the key to stop
at your desired set-point.
h
A buzzer sounds if the meter reads a %MC higher than the
set-point.
4