Setup menu – DeFelsko PosiTector 6000 v.6.2 for Basic (1) and Standard (2) models User Manual
Page 11

11
-zeros all on-screen
Statistics and HiLo tabulations.
is above the
Hi limit. The LED will blink red if
readings are outside limits. Press (+) to clear
HiLo
readings.
Reset restores factory settings and returns
the Gage to a known, out-of-the-box
condition. It is handy when settings have been
changed, if the Gage behaves unusually, or if a
calibration adjustment is not possible for some
reason. The following occurs:
- all stored measurements are erased.
- calibration adjustments are cleared and returned
to the Gage’s factory calibration settings.
- This symbol appears on the display: It
disappears if a calibration adjustment is made by
the user.
- menu setting are returned to the following:
Hi Res = OFF Memory = OFF
Cal Lock = OFF Statistics = OFF
Hi Lo Alarm = OFF N Lock = OFF (FN models only)
A more thorough
Reset can be performed when the
Gage is powered down by holding the (+) button
until the Reset symbol appears. This is handy
when the Gage fails to power-up or operate properly.
It performs the same function as a menu Reset with
addition of
Units = microns, Flip Display = Normal and
Language = English.
-Keep the Gage away from metal during a
Reset.
Setup Menu
Setup Menu
NOTE: