Under a pressurized system removal of the sensor with-
out the installation of isolation valves can be dangerous.

If in-situ installation is required, bypass mounting is prefer-
able; make sure to install the sensor at the upper surface of
the gas line to minimize its exposure to liquid water, should
condensation occur, the XTR65W sensor is best suited for
these applications. Also consider the need to isolate (depres-
surize) before installing or removing the sensor.