What to do if – Seca 763 User Manual

Page 25

background image

Model 763

25

7. What to do if …

… no weight or height readout appears?
– Is the device switched on?
– Is the plug between the power supply connector in the rod and base properly connect-

ed?

– Is the plug contact of the AC adaptor properly connected on the device rod?
– Is the plug contact for connecting the display and the base properly connected?

0.0

does not appear before weighing?

– Press the start button again – there must not be any load on the scales – and only its

feet should be in contact with the floor.

… one segment is illuminated constantly or not at all?
– The relevant segment has a fault. Inform Servicing.

St0P

appears in the weight display?

– Maximum load has been exceeded.

… the weight display is blinking?
– If you have just activated the Hold or Tare function, please wait until the scales have

detected a steady weight value. The display will then stop flashing.

– If there has been no activation of a function beforehand, remove the load from the

scales and wait until

0.0

is displayed, then weigh again.

… the display

appears?

– The ambient temperature of the device is too high or too low. Place the device where

the ambient temperature is between +10 ˚C and +40 ˚C. Wait about 15 minutes until

the device has adjusted to the ambient temperature and weigh again.

… the display and a number appear?
– Use the Start button to switch off the device and start again. After this the device will

function normally.

– If this is not the case, disconnect the power supply by briefly unplugging the AC adap-

tor. If this measure is equally unsuccessful, inform the Service department.

appears in the height display?

– Slide the head slide up or down quite a ways. This will make the head slide pass a mark

which can be evaluated and the system can now calculate the position correctly again.