Calibration Confirm:
Press the button to confirm the calibration value after calibration

is completed. Instrument will return to normal operation mode.

Note: In the self-check and calibration mode, self-calibration

coefficient is from 1600-2495. If user calibrates in this range,
and accuracy is still not achievable, please contact an Authorized
Service Center or Dealer for repair.

• After Calibration is complete, user must turn the instrument

off and then back on. Calibration is then complete.

• When calibrating, the user must check accuracy of both the

X and Y Axis, until both are within accuracy specification.