8. Self-Check & Fine Calibration

IMPORTANT: It is the responsibility of the user to verify the
calibration of the instrument before each use.

• Find a road, parking lot, or level plane (slope should be less than 4”

at 100’/30m).

• Pick 2 points (A and B) whose distance is about 100’ or 30m

between each other. The distance does not need to be measured
exactly, mark the 2 points.

• Put the pipe laser behind point A, turn it on, and let it warm-up for

10 minutes.

• Set the grade to be 00.000%.

• Point the laser beam to go through points A and B, make sure the

laser is in its self-leveling range.

• Accurately measure the height from the center of the laser beam

point A and point B, note it as A1 and B1.

• Put the laser behind point B and let the laser beam go through

points A and B making sure the laser is in its self-leveling range.