Page 4-13 “clock display, Page 13-5 “error codes, Page 13-8 “calibrations – Yokogawa DR240 User Manual
Page 4: Page 14-3 “standard computation functions, Scaling

IM DR232-01E-9A 4/5
for IM DR232-01E 9th Edition
Page 4-13 “Clock Display”
The date and time can be displayed on sub-display 2.
According to the set time in 3.10, “Setting the Date and Time” (see page 3-35), the current date and
time is displayed.
The display shows the month, day, year and hour, minute, second in this sequence.
Page 13-5 “Error Codes”
The following error codes have been added to the list.
Error Codes
Error
Corrective Action
012
None of the channels specified in execution of
initial balance were strain input channels.
Enter a correct channel.
013
Attempted “REP RECALL START” without hourly,
daily, or monthly report data.
032
Set to contiguous channels on the power monitor
module.
Enter a correct channel.
047
Entered a wiring method for which there is no
setting for the power monitor module.
Enter the correct wiring method.
107
Attempted to change ranges or time while report
was starting.
Do not make changes.
131
Media write error
Exchange the medium.
137
Attempted to start computation or execute a
procedure during saving of settings to a medium
or while reading from a medium.
Start the procedure or computation after
completion of the media operation.
138
Media drive error
Exchange the medium. If the error occurs again
after exchanging the medium, servicing is
required.
146
An initial balancing error occurred during
calibration.
Check the wiring and try calibration again.
Page 13-8 “calibrations”
For “Strain measurement” under “Connection,” the wiring diagram for the bridge box used
for the DU500-14 has been changed to the following:
Connect CH2 only
Switch the
bridge head
R2
R4
R1
R3
ON
1 2 3 4 5
1
2
3
4
5
6
7
8
OFF
SW
Bridge head
(701955 or 701956)
SW1
OFF
SW2
OFF
SW3
OFF
SW4
ON
SW5
OFF
Page 14-3 “Standard Computation Functions”
Scaling
Measurement accuracy for scaling: measurement accuracy for scaling (digits) = measurement
accuracy (digits) × scaling span (digits) / measurement span (digits) + 2 digits. Numbers below
the decimal point are rounded up.
±(((0.05/100) Ч 5000) + 2) Ч (2000/4000) + 2 = ±4.25
Measurement accuracy = ±5 digits = ±0.005 V