Page 13-5 Added to the “13.4 Error Codes”

The following error codes have been added to the list.

Error Codes

Error

Corrective Action

013

Attempted “REP RECALL START” without hourly,

daily, or monthly report data.

032

Set to contiguous channels on the power monitor

module.

Enter a correct channel.

047

Entered a wiring method for which there is no

setting for the power monitor module.

Enter the correct wiring method.

107

Attempted to change ranges or time while report

was starting.

Do not make changes.

131

Media write error

Exchange the medium.

137

Attempted to start computation or execute a

procedure during saving of settings to a medium

or while reading from a medium.

Start the procedure or computation after completion of

the media operation.

138

Media drive error

Exchange the medium. If the error occurs again after

exchanging the medium, servicing is required.

170

The IP address does not belong to any of the

classes, A, B, or C.

Set the correct IP address.

171

There is a hole in the mask or the host address

section is not released.

Set the correct subnet mask.

172

The net address section including the subnet does

not match the subnet section of the IP address.

Set the correct net address.

173

The host address section of the IP address is

either all zeroes or all ones.

Set the correct IP address.

Page 14-9 “Standard Computation Functions”

Scaling

Measurement accuracy for scaling: measurement accuracy for scaling (digits) = measurement
accuracy (digits) × scaling span (digits) / measurement span (digits) + 2 digits. Numbers below
the decimal point are rounded up.

±(((0.05/100) Ч 5000) + 2) Ч (2000/4000) + 2 = ±4.25
Measurement accuracy = ±5 digits = ±0.005 V

Page 14-14 “Normal Operation Conditions”

Installation category based on IEC61010-1, DSA22.2 No 61010-1.
II

*1

Pollution degree based on IEC61010-1, DSA22.2 No 61010-1.
2

*2

Warm-up time
At least 30 minutes after power switch-on.

*1

Describes a number which defines a transient overvoltage condition. It implies the regulation for impulse

withstand voltage. “II” applies to electrical equipment which is supplied from fixed installations like distribution
boards.

*2

Describes the degree to which a solid, liquid, or gas which deteriorates dielectric strength or surface

resistivity is adhering. “2” applies to normal indoor atmosphere. Normally, only non-conductive pollution
occurs.

Page 14-15 “EMC Conformity Standards”

Please refer to these specifications instead of the one printed in the user’s manual.

Safety and EMC Standards

CSA

CSA22.2 No.61010-1, installation category II, pollution degree 2

UL

UL61010-1 (CSA NRTL/C)

C-Tick

EN55011 compliance, Class A, Group 1

KC marking

Electromagnetic wave interference prevention standard, electromagnetic wave protection

standard compliance