Page 14-9 “standard computation functions, Page 14-14 “normal operation conditions, Page 14-15 “emc conformity standards – Yokogawa DR240 User Manual
Page 3

IM DR231-01E-98 3/3
for IM DR231-01E 9th Edition
Page 13-5 Added to the “13.4 Error Codes”
The following error codes have been added to the list.
Error Codes
Error
Corrective Action
013
Attempted “REP RECALL START” without hourly,
daily, or monthly report data.
032
Set to contiguous channels on the power monitor
module.
Enter a correct channel.
047
Entered a wiring method for which there is no
setting for the power monitor module.
Enter the correct wiring method.
107
Attempted to change ranges or time while report
was starting.
Do not make changes.
131
Media write error
Exchange the medium.
137
Attempted to start computation or execute a
procedure during saving of settings to a medium
or while reading from a medium.
Start the procedure or computation after completion of
the media operation.
138
Media drive error
Exchange the medium. If the error occurs again after
exchanging the medium, servicing is required.
170
The IP address does not belong to any of the
classes, A, B, or C.
Set the correct IP address.
171
There is a hole in the mask or the host address
section is not released.
Set the correct subnet mask.
172
The net address section including the subnet does
not match the subnet section of the IP address.
Set the correct net address.
173
The host address section of the IP address is
either all zeroes or all ones.
Set the correct IP address.
Page 14-9 “Standard Computation Functions”
Scaling
Measurement accuracy for scaling: measurement accuracy for scaling (digits) = measurement
accuracy (digits) × scaling span (digits) / measurement span (digits) + 2 digits. Numbers below
the decimal point are rounded up.
±(((0.05/100) Ч 5000) + 2) Ч (2000/4000) + 2 = ±4.25
Measurement accuracy = ±5 digits = ±0.005 V
Page 14-14 “Normal Operation Conditions”
Installation category based on IEC61010-1, DSA22.2 No 61010-1.
II
*1
Pollution degree based on IEC61010-1, DSA22.2 No 61010-1.
2
*2
Warm-up time
At least 30 minutes after power switch-on.
*1
Describes a number which defines a transient overvoltage condition. It implies the regulation for impulse
withstand voltage. “II” applies to electrical equipment which is supplied from fixed installations like distribution
boards.
*2
Describes the degree to which a solid, liquid, or gas which deteriorates dielectric strength or surface
resistivity is adhering. “2” applies to normal indoor atmosphere. Normally, only non-conductive pollution
occurs.
Page 14-15 “EMC Conformity Standards”
Please refer to these specifications instead of the one printed in the user’s manual.
Safety and EMC Standards
CSA
CSA22.2 No.61010-1, installation category II, pollution degree 2
UL
UL61010-1 (CSA NRTL/C)
C-Tick
EN55011 compliance, Class A, Group 1
KC marking
Electromagnetic wave interference prevention standard, electromagnetic wave protection
standard compliance