beautypg.com

Yokogawa Value Series FX1000 User Manual

Introduction, Safety precautions, Installation

background image

IM 04L21B01-03EN page 1/2

Introduction

Thank you for purchasing the FX1000 (FX). This manual describes the safty

precautions and installation and wiring procedures of the FX1000.

For FX1000 topics such as the package contents, safety precautions, and

detailed handling procedures, see the electronic manuals, which are stored in

PDF format on the included CD.

1. Safety Precautions

The following safety symbols are used on the product and in this manual.

WARNING

Calls attention to actions or conditions that could cause serious

or fatal injury to the user, and indicates precautions that should be

taken to prevent such occurrences.

CAUTION

Calls attention to actions or conditions that could cause injury to

the user or damage to the instrument or property and indicates pre-

cautions that should be taken to prevent such occurrences.

Protective ground terminal

Alternating current

Direct current

Note

Identifies important information required to operate the instrument.

n

Safety Precautions

• This instrument conforms to IEC safety class I (provided with terminal for protec-

tive grounding), Installation Category II, and EN61326-1 (EMC standard), Mea-

surement Category II (CAT II)*.

* Measurement category II (CAT II) applies to measuring circuits connected to low voltage

installation, and electrical instruments supplied with power from fixed equipment such as

electric switchboards.

• This instrument is an EN61326-1 (EMC standard) class A instrument (for use in

commercial, industrial, or business environments).

• The general safety precautions described here must be observed during all phases

of operation. If the FX is used in a manner not described in this manual, the FX

safety features may be impaired. Yokogawa Electric Corporation assumes no li-

ability for the customer’s failure to comply with these requirements.

• The FX is designed for indoor use.

n

About This Manual

• Please pass this manual to the end user. We also ask you to store this manual in a safe place.

• Read this manual thoroughly and have a clear understanding of the product before operation.

• This manual explains the functions of the product. It does not guarantee that the

product will suit a particular purpose of the user.

n

Precautions Related to the Protection, Safety, and Alteration of the Product

• For the protection and safe use of the product and the system in which this product

is incorporated, be sure to follow the instructions and precautions on safety that are

stated in this manual whenever you handle the product. Take special note that if you

handle the product in a manner that violates these instructions, the protection function-

ality of the product may be damaged or impaired. In such cases, YOKOGAWA does not

guarantee the quality, performance, function, and safety of product.

• When installing protection and/or safety circuits such as lightning protection devices

and equipment for the product and control system or designing or installing separate

protection and/or safety circuits for fool-proof design and fail-safe design of the pro-

cesses and lines that use the product and the control system, the user should imple-

ment these using additional devices and equipment.

• If you are replacing parts or consumable items of the product, make sure to use parts

specified by YOKOGAWA.

• This product is not designed or manufactured to be used in critical applications that di-

rectly affect or threaten human lives. Such applications include nuclear power equipment,

devices using radioactivity, railway facilities, aviation equipment, air navigation facilities,

aviation facilities, and medical equipment. If so used, it is the user’s responsibility to in-

clude in the system additional equipment and devices that ensure personnel safety.

• Do not modify this product.

WARNING

l

Use the Correct Power Supply

Ensure that the source voltage matches the voltage of the power

supply before turning ON the power.

l

Connect the Protective Grounding Terminal

Make sure to connect the protective grounding to prevent

electric shock before turning ON the power.

l

Do Not Impair the Protective Grounding

Never cut off the internal or external protective grounding wire

or disconnect the wiring of the protective grounding terminal.

Doing so invalidates the protective functions of the instrument

and poses a potential shock hazard.

l

Do Not Operate with Defective Protective Grounding

Do not operate the instrument if the protective grounding might

be defective. Also, make sure to check them before operation.

l

Do Not Operate in an Explosive Atmosphere

Do not operate the instrument in the presence of flammable

liquids or vapors. Operation in such an environment constitutes

a safety hazard. Prolonged use in a highly dense corrosive gas

(H

2

S, SOx, etc.) will cause a malfunction.

l

Do Not Remove Covers

The cover should be removed by YOKOGAWA’s qualified

personnel only. Opening the cover is dangerous, because some

areas inside the instrument have high voltages.

l

Ground the Instrument before Making External Connections

Connect the protective grounding before connecting to the item

under measurement or control unit.

l

Damage to the Protection

Operating the instrument in a manner not described in this

manual may damage the instrument’s protection.

CAUTION

This instrument is a Class A product. Operation of this instrument in a

residential area may cause radio interference, in which case the user is

required to take appropriate measures to correct the interference.

n

Exemption from Responsibility

• YOKOGAWA makes no warranties regarding the product except those stated in

the WARRANTY that is provided separately.

• YOKOGAWA assumes no liability to any party for any loss or damage, direct or

indirect, caused by the user or any unpredictable defect of the product.

2. Installation

n

Installation Location

Install the FX indoors in an environment that meets the following conditions:

• Instrumentation Panel

The FX is designed to be installed in an instrumentation panel.

• Well-Ventilated Location

To prevent overheating, install the FX in a well-ventilated location. For the panel

cut dimensions when arranging multiple FXs, see “External Dimensions and Panel

Cutout Dimensions.” When other instruments are installed next to the FX, follow

the panel cut dimensions to provide adequate space around the FX.

• Minimal Mechanical Vibrations

Install the FX in a location that has minimal mechanical vibrations. Installing the FX

in a location that is subject to large levels of mechanical vibration will not only put

added stress on its components, it may also impede ordinary measurement.

• Level Location

Install the FX in a level location so that it is not slanted to the left or the right (how-

ever, the FX can be inclined up to 30 degrees backward for panel mounting).

Note

Condensation may form when moving the FX from an environment whose temperature or

humidity is low to an environment whose temperature or humidity is high, or when there

is a sudden change in temperature. Temperature or humidity changes may also result in

thermocouple measurement errors. In these kinds of circumstances, let the FX adjust to

the new environment for at least an hour before using it.

Do not install the FX in the following places.

• Outdoors

• In Direct Sunlight or Near Heat Sources

Install the FX in a place that is near room temperature (23°C) and that is not sub-

ject to large temperature fluctuations. Placing the FX in direct sunlight or near heat

sources can cause adverse effects on the internal circuitry.

• Where an Excessive Amount of Soot, Steam, Moisture, Dust, or Corrosive

Gases Are Present

Soot, steam, moisture, dust, and corrosive gases will adversely affect the FX.

Avoid installing the FX in such locations.

• Near Strong Magnetic Field Sources

Do not bring magnets or instruments that produce electromagnetic fields close to the

FX. Operating the FX near strong magnetic fields can cause measurement errors.

• Where the Display Is Difficult to See

The FX uses an LCD screen, so it is difficult to view the display from an extreme

angle. Install the FX so that the user can view the display directly from the front.

n

Installation Procedure

Use a steel panel that is 2 mm to 26 mm thick.

1) Insert the FX through the front of the panel.

2) Mount the FX to the panel using the included mounting brackets as shown in the figure below.

• Use two mounting brackets to support the top and bottom or the left and right sides of the

case (remove the stickers that are covering the holes before you attach the brackets).

• Follow the procedure below to mount the FX to the panel.

• First, attach the two mounting brackets and temporarily tighten the mounting screws.

• Next, fix the FX in place by tightening the mounting screws with the appropriate torque.

When the FX is approximately perpendicular to the panel, press the mounting brackets

so that they are in contact with the case, and fully tighten the mounting screws.

• Tighten the mounting bracket screws until you hear clicks.

CAUTION

l

Using more than the appropriate torque to tighten the screws

can deform the case or damage the brackets.

l

Be sure not to insert foreign objects or tools into the case

through the mounting bracket holes.

l

Do not touch the screw at the top of the mounting bracket

(A). Loosening or tightening this screw may cause the FX to

malfunction.

Front

Panel

Panel

Mounting bracket

Mounting bracket

Torque driver

(Phillips or flat blade)

Mounting screw

FX case

Attach firmly

(In the figure, the mounting brackets are used on the top and bottom of the case.)

Mounting screw

Screws temporarily fastened

Fixed

Rubber packing for dust

and water protection

(except side-by-side

mounting)

A

Note

To achieve sufficient dust proofing and waterproofing, mount the FX in the middle of

the panel cut out. However, do not use the rubber packing if you are mounting two

instruments side by side or one on top of the other.

n

External Dimensions and Panel Cutout Dimensions

136.5

+0.4 0

Unit: mm (approx. inch)

If not specified, the tolerance is ±3%.

However, in cases of less than 10mm, the tolerance is ±0.3 mm.

Panel thickness

(Dimensions after attaching

the mounting bracket)

If not installing the rubber packing for dust and water protection,

*1: 20.2 (0.80), *2: 163.7 (6.44), *3: 110.0 (4.33).

(*1)

(*2)

(*3)

(Dimensions after attaching

the mounting bracket)

136.5 (5.37)

144.0 (5.67)

161.7 (6.37)

72.0 (2.83)

72.0 (2.83)

12.6 (0.50)

12.6 (0.50)

±0.3(0.01)

(2.74)

74.5 (2.93)

161.7 (6.37)

22.2 (0.87) 161.7 (6.37)

108.0 (4.25)

144.0 (5.67)

2 to 26

(0.08 to 1.02)

Mounting bracket

69.5

Panel cut dimensions

MIN 175

(6.89)

L

+ 2 (0.08)

0

137 (5.39)

+ 2 (0.08)

0

Side-by side mounting

(vertically; max. 3 units)

137

(5.39)

MIN 175

(6.89)

0

L

+ 2 (0.08)

0

Side-by-side mounting

(horizontally)

137 (5.39)

+ 2 (0.08)

0

137

(5.39)

+ 2 (0.08)

0

Single-unit mounting

L

+ 2 (0.08)

0

2

3

4

5

6

7

8

9

10

n

282 (11.10)

426 (16.77)

in mm (approx. inches)

570 (22.44)

714 (28.11)

858 (33.78)

1434 (56.46)

1290 (50.79)

1146 (45.12)

1002 (39.45)

(144 × n) – 6 [(5.67 × n) – 0.24]

Units

+ 2 (0.08)

External dimensions

3. Wiring

n

Input Signal Wiring

WARNING

To prevent electric shock while wiring, make sure that the power

supply is turned off.

CAUTION

l

Applying a strong tension to the input and output signal cables

connected to the FX may damage the cables or the FX terminals.

To avoid applying tension directly to the terminals, fix all cables

to the rear of the mounting panel.

l

To prevent fire, use signal cables with a temperature rating of 70°C

or more.

l

Do not apply voltages that exceed the following values to the

input terminals. Doing so may damage the FX.

• Maximum input voltage: ±60 VDC

• Maximum common mode voltage: ±60 VDC (under

measurement category II conditions)

l

The FX is an installation category II product.

Precautions to Be Taken While Wiring

Take the following precautions when wiring the input signal cables.

When using a screw terminal, we recommend that you use a crimp-on lug with an

insulation sleeve (designed for 3 mm screws).

Take measures to prevent noise from entering the measurement circuit.

• Move the measurement circuit away from the power cable (power circuit) and ground circuit.

• Ideally, the object being measured should not generate noise. However, if this is

unavoidable, isolate the measurement circuit from the object. Also, ground the ob-

ject being measured.

• Shielded wires should be used to minimize the noise caused by electrostatic induc-

tion. Connect the shield to the ground terminal of the FX as necessary (make sure

you are not grounding at two points).

• To minimize noise caused by electromagnetic induction, twist the measurement

circuit wires at short, equal intervals.

• Make sure to earth ground the protective ground terminal through minimum resis-

tance (less than 100 Ω).

When using internal reference junction compensation on the thermocouple input,

take measures to stabilize the temperature at the input terminal.

• Always use the terminal cover.

• Do not use thick wires which may cause large heat dissipation (we recommend a

cross sectional area of 0.5 mm

2

or less).

• Make sure that the ambient temperature remains reasonably stable. Large tem-

perature fluctuations can occur if a nearby fan turns on or off.

Connecting the input wires in parallel with other devices can cause signal degradation,

affecting all connected devices. If you need to make a parallel connection, then

• Turn the burnout detection function off.

• Ground the instruments to the same point.

• Do not turn other instruments on or off during operation. This can have adverse ef-

fects on the other instruments.

• RTDs cannot be wired in parallel.

Input terminal block

of the FX1002

Input terminal block

of the FX1004

Input terminal block

of the FX1006

Input terminal block

of the FX1008

Input terminal block

of the FX1010

Input terminal block

of the FX1012

For TC input, use shielded compensating lead wires for wiring.

For RTD input, lead wire resistance per wire of 10 Ω or less. Make the resistances of

the three wires equal.

For DCA input, example: for 4 to 20 mA input, use a shunt resistor of 250 Ω ± 0.1%.

+/A

/b

CH1

–/B

CH2

+/A

/b

CH1

–/B

CH2

CH3

CH4

+/A

/b

CH1

–/B

CH2

CH3

CH4

CH5

CH6

+/A

/b

CH1

CH7

–/B

CH2

CH8

CH3

CH4

CH5

CH6

+/A

/b

CH1

CH7

–/B

CH2

CH8

CH9

CH10

CH3

CH4

CH5

CH6

+/A

/b

CH1

CH7

–/B

CH2

CH8

CH9

CH10

CH11

CH12

CH3

CH4

CH5

CH6

Note

RTD input terminals A and B are isolated on each channel. Terminal b is shorted

internally across all channels. However, terminal b is also isolated on each channel

on models with the /N2 option (3 leg isolated RTD).

Installation

Guide

IM 04L21B01-03EN

FX1000

Safety Precautions and

Installation Guide

2nd Edition : Sept. 2012