Wiring – Yokogawa FU20/PH20 All-in-One pH/ORP Sensor User Manual
Page 16

IM 12B6J3-E-H
12
TEMPERATURE
TEMPERATURE
REFERENCE
REFERENCE SCREEN
SOLUTION GROUND
Pt (ORP Measure)
GLASS SCREEN
(11)
(12)
(13)
(17)
(14)
(15)
(16)
16
15
11
12
13
14
(Not used)
T2
T1
RE
SE
GE
S
ORP WIRING DIAGRAM with normal reference
TEMPERATURE
TEMPERATURE
REFERENCE (pH)
REFERENCE SCREEN
(SOLUTION GROUND)
Pt (ORP Measure)
GLASS SCREEN
(11)
(12)
(13)
(17)
(14)
(15)
(16)
13
11
12
15
14
(Not used)
16
T2
T1
RE
SE
GE
S
ORP WIRING DIAGRAM with pH sensor as reference
TEMPERATURE
TEMPERATURE
REFERENCE
REFERENCE SCREEN
SOLUTION GROUND
GLASS (Measure)
GLASS SCREEN
(11)
(12)
(13)
(17)
(14)
(15)
(16)
11
12
13
14
16
T2
T1
RE
SE
GE
S
15
pH (& ORP) WIRING DIAGRAM
6. WIRING
6-1. Conventional pH (& ORP) Wiring
Connect the PH20 or FU20 to the EXA or
EXAxt PH Analyzer as shown (Fig 12). With
this configuration, it is possible to measure
ORP (or rH) at the same time (Refer to the
EXA or EXAxt Manual for appropriate impe-
dance jumper and Service Code settings).
6-2. Wiring for ORP measurement with
normal reference
Connect the PH20 or FU20 to the EXA or
EXAxt PH Analyzer as shown (Fig 13).
Refer to the EXA or EXAxt Manual for
appropriate impedance jumper and Service
Code settings).
6-3. Wiring for ORP measurement with
pH reference
Connect the PH20 or FU20 to the EXA or
EXAxt PH Analyzer as shown (Fig 14).
Refer to the EXA or EXAxt Manual for
appropriate impedance jumper and Service
Code settings).
Figure 12.
Figure 13.
Figure 14.