IM 1E8B0-01E

13-2

(4) Maintenance and Repair

The instrument modification or parts replacement by other than authorized
representative of Yokogawa Electric Corporation is prohibited and will void
the approval of Factory Mutual Research Corporation.

HAZARDOUS LOCATIONS

Magnetic Flow Converter

Conduit

Sealing Fitting

18"(457mm) Max.

13.2 CSA (Only for sizes 15 to 100 mm(0.5 to 4 in))

(1) Technical Data

Optional code /CF11
Applicable Standard:

CSA Standard C22.2
No.0, No.0.4, No.0.5, No.25, No.30, No.94, No.61010-1

Certificate:

1500865

Explosion proof for Class I, Groups B, C and D; Class II, Groups E, F and G; Class III
Temp. Code: T6 T5 T4
Process Temp.: 70 85 120

°C

Ambient Temp.: –20

°C to +50°C (–4 to 122°F)

Maximum power supply voltage: 240 Vac/ 120 Vdc
Enclosure: TYPE 4X

(2) Installation

All wiring shall comply with Canadian Electrical Code Part I and Local
Electrical Codes.
In hazardous location, wiring shall be in conduit as shown in the figure.

CAUTION :

SEAL ALL CONDUITS WITHN 50cm OF THE ENCLOSURE’

UN SCELLEMENT DOIT ÉTRE INSTALLÉ À MOINS DE

50cm DU BOÎTIER.