Checking capacity loss – ZOLL X Series Monitor Defibrillator Rev A User Manual
Page 8

U
SING
THE
SurePower II Battery Pack
6
www.zoll.com
9650-000840-01 Rev. A
Extended Storage
Do not store the SurePower II Battery Pack for extended periods in a fully charged state. If you 
must store the battery pack for an extended period 
1. Charge the SurePower II Battery Pack until three runtime LEDs light (this takes
approximately 1.5 hours when charging a fully depleted battery).
2.
Recharge the SurePower II Battery Pack until three runtime LEDs light every 4 months.
Checking Capacity Loss
To check battery capacity (the amount of charge the battery can hold), press and hold down the 
Display button for at least 5 seconds. 
Press and hold down the Display button for at least 
5 seconds. 
While you hold down the
Display
button, the
runtime indicator displays the battery runtime. 
When you release the 
Display
button, the
runtime indicator displays the battery capacity 
for 2 seconds, then redisplays the battery 
runtime.
During the 2 seconds after you release the
Display
button,
the runtime indicator displays
battery capacity and will light five LEDs to 
indicate that the capacity is equal or close to 
that of a new battery. 
After displaying battery capacity for 2 seconds, 
all LEDs blink and then display the battery’s 
runtime.
Over time, as you use a SurePower II Battery 
Pack
, the battery capacity decreases and this
capacity check will light fewer LEDs. When the 
capacity check lights only a single LED, the 
battery is at or near 60% of its original capacity, 
and you should remove the battery from service.
If the Recalibration or Fault LED lights after 
you release the Display button, the runtime 
indicator will not display the capacity or run time 
for that battery.