NTRODUCTION
2
www.zoll.com
9650-0809-01 Rev. A
•
Do not place foreign objects into battery sockets.
•
Do not use the Propaq MD Battery Pack in the presence of flammable agents (such as
gasoline), oxygen-rich atmospheres, or flammable anesthetics. Using the device in the
presence of flammable agents can cause an explosion.
•
Do not immerse or set the Propaq MD Battery Pack in liquid. Using the battery near or
within puddles of water may present a shock hazard to the operator, patient, and nearby
personnel.
•
Use only ZOLL chargers to charge ZOLL batteries.
•
If the display indicators on the Propaq MD Battery Pack are not consistent with the
indicators on the SurePower Charger Station, a fault condition may be present. Remove the
battery from service as soon as practical, and verify the correct operation of the charger,
defibrillator, and battery.
Cautions
•
Follow all recommended maintenance instructions. If a problem occurs, obtain service
immediately.
•
Do not use alcohol or ketones (MEK, acetone, etc.) to clean Propaq MD Battery Packs.
•
Do not sterilize or autoclave Propaq MD Battery Packs.
•
Avoid storage of fully depleted Propaq Battery Packs.
•
Do not install batteries into a Propaq MD unit that is not plugged into live AC mains when
storage may exceed 90 days. Battery damage may occur.
•
If you are disposing of a Propaq MD Battery Pack, contact your local authorities to
determine the requirements for the recycling and disposal of electrical equipment and follow
those requirements.
Notes
•
Propaq MD Battery Packs may feel warm to the touch following a large number of shocks
delivered at maximum energy or after recharging.
•
You can get additional information about the Propaq MD Battery Pack and its use from the
Propaq MD Operator’s Manual (for example, labeling information -- symbols used, Serial
Number format, Date of Manufacture, and so on).