beautypg.com

Cautions – ZOLL Propaq MD Rev A User Manual

Page 4

background image

I

NTRODUCTION

2

www.zoll.com

9650-0809-01 Rev. A

Do not place foreign objects into battery sockets.

Do not use the Propaq MD Battery Pack in the presence of flammable agents (such as

gasoline), oxygen-rich atmospheres, or flammable anesthetics. Using the device in the

presence of flammable agents can cause an explosion.

Do not immerse or set the Propaq MD Battery Pack in liquid. Using the battery near or

within puddles of water may present a shock hazard to the operator, patient, and nearby

personnel.

Use only ZOLL chargers to charge ZOLL batteries.

If the display indicators on the Propaq MD Battery Pack are not consistent with the

indicators on the SurePower Charger Station, a fault condition may be present. Remove the

battery from service as soon as practical, and verify the correct operation of the charger,

defibrillator, and battery.

Cautions

Follow all recommended maintenance instructions. If a problem occurs, obtain service

immediately.

Do not use alcohol or ketones (MEK, acetone, etc.) to clean Propaq MD Battery Packs.

Do not sterilize or autoclave Propaq MD Battery Packs.

Avoid storage of fully depleted Propaq Battery Packs.

Do not install batteries into a Propaq MD unit that is not plugged into live AC mains when

storage may exceed 90 days. Battery damage may occur.

If you are disposing of a Propaq MD Battery Pack, contact your local authorities to

determine the requirements for the recycling and disposal of electrical equipment and follow

those requirements.

Notes

Propaq MD Battery Packs may feel warm to the touch following a large number of shocks

delivered at maximum energy or after recharging.

You can get additional information about the Propaq MD Battery Pack and its use from the

Propaq MD Operator’s Manual (for example, labeling information -- symbols used, Serial

Number format, Date of Manufacture, and so on).

This manual is related to the following products: