Additional messages and troubleshooting, Additional specifications and changes – ZOLL M Series Defibrillator Rev E BiPhasic User Manual
Page 6

OPTION INSERT
9650-0209-01
Biphasic- 6
Additional Messages and TroubleShooting
The following chart lists the messages that may appear on the M Series unit relating to the Biphasic Option, why the message
appeared, and the action(s) to take.
The operator should become thoroughly familiar with this information before using the device.
Message
Possible
Cause(s)
Recommended
Action(s)
200 J MAX BIPHASIC
Appears when trying to select an energy higher
No higher energy is available.
than
200
Joules.
Use
200J
setting.
BRIDGE TEST FAILED
Biphasic module not operating properly while
Try charging again. Attempt to
charging.
clear the message by turning the
Selector
Switch
to
OFF
then
back
to
the
desired
operating
mode.
If
Fault
persists
contact
ZOLL
Technical
Service
Department.
BRIDGE SHORT
Current higher than expected was detected.
Ensure pads/paddles are used
properly.
Attempt
to
clear
the
message
by
turning
the
Selector
Switch to OFF then back to the
desired
operating
mode.
If
Fault
persists
contact
ZOLL
Technical
Service
Department.
Additional Specifications and Changes
General
Refer to M Series Operator’s Guide for ALL Specifications except the following:
Waveform:
Rectilinear Biphasic
Energy Settings:
1, 2, 3, 4, 5, 6, 7, 8, 9, 10, 15, 20, 30, 50, 75, 100, 120, 150, 200
Charge Time:
Less than 6 seconds with a new fully charged battery
(first 15 charges at 200J)
Messages:
“200 J MAX BIPHASIC”, “BRIDGE TEST FAILED”, “BRIDGE SHORT”
Operating
Time:
For a new, fully charged battery pack at 20°C: 40 defibrillator
discharges
at
maximum
energy (200J), or 2.75 hours minimum of
continuous
ECG
monitoring, or 2.25 hours of continuous ECG
monitoring/pacing at 60 mA, 80 beats/min.