2
ZOLL M Series XL Battery
General Information
Defibrillator Compatibility
M Series units must have software version 30.00 or higher and be outfitted with the appropriate
handle assembly. New M Series units may be ordered with the XL Battery option. Older M Series
units may be upgraded if the current software version is 20.00 or higher. All M Series CCT units are
XL Battery-ready.
The XL Battery is not intended for use in ZOLL 900/1200/1400/1600/1700/2000 series units.
Charger Compatibility
The XL Battery is only compatible with ZOLL Base PowerCharger
4x4
or Base PowerCharger
1x1
units
labeled “XL Battery Ready.”
Older ZOLL Base PowerCharger
4x4
or Base PowerCharger
1x1
units may be upgraded.
All XL Battery ready chargers are backward compatible with the PD4410 battery pack.
Battery Maintenance
Test batteries regularly. At a minimum, ZOLL recommends testing batteries every three months. For
additional information on battery maintenance, refer to the ZOLL document, Battery Management
Program (#9650-0019).
Environmental Specifications
Operating Note
The XL Battery may feel warm to the touch following an excessive number of defibrillation
discharges at maximum energy.
Charge Time
The XL battery charge times are as follows:
Cleaning
Clean the ZOLL XL Battery with a soft damp cloth and mild soap.
CAUTION! Do not immerse or set battery in fluids; do not use alcohol, ketones
(MEK, acetone, etc.); and do not autoclave.
Temperature
0°C to 50°C (operating)
15°C to 35°C (optimal battery charging)
-40°C to 70°C (storage)
Humidity
5% to 95% (relative humidity, non-condensing)
Charging Device
Maximum Charge Time
M Series - unit plugged in and OFF
7.2 hours
M Series - unit plugged in and in use
43 hours
SurePower Charger Station Quick Charge
6 hours
SurePower Charger Station Auto Test
12 hours