Guidance and manufacturer’s declaration, Electromagnetic emissions – ZOLL E Series Monitor Defibrillator Rev R User Manual
Page 105

9650-1210-01 Rev. R
A-7
Guidance and Manufacturer’s Declaration
—
Electromagnetic Emissions
The E Series unit is intended for use in the electromagnetic environment specified below. Ensure that the E Series unit
is used in such an environment.
Emissions Test
Compliance
Electromagnetic Environment — Guidance
RF emissions
CISPR11
Group 1
The E Series unit uses RF energy only for its internal func-
tion. Therefore, its RF emissions are very low and are not
likely to cause any interference in nearby electronic equip-
ment.
RF emissions
CISPR 11
Class B
Class A
The E Series unit is suitable for use in all establishments,
including domestic establishments and those directly con-
nected to the public low-voltage power supply network that
supplies buildings used for domestic purposes.
Class A equipment is equipment suitable for use in all estab-
lishments other than domestic and those directly connected
to a low voltage power supply network which supplies build-
ings used for domestic purposes.
Harmonic Emission
IEC 61000-3-2
Class A
Voltage Fluctuations/
Flicker Emission
IEC 61000-3-3
Complies
Medical Electrical Equipment needs special precautions regarding EMC and needs to be installed and put into ser-
vice according to EMC information provided in this document.