ZOLL E Series Monitor Defibrillator Rev C User Manual
Page 15

Troubleshooting
9650-1215-01 Rev. C
13
FAX ERROR message is displayed.
•
A dialing problem other than a busy line, no carrier, or no
connection occurred.
•
The E Series automatically re-transmits until the transmission is
successful or the E Series operator presses the Abort softkey.
FAX HANGUP message is displayed.
•
Receiving phone is not receiving transmission. The E Series
automatically attempts to transmit until the transmission is
successful or the E Series operator presses the Abort softkey.
FAX INTERRUPTED message is displayed.
•
Power cycle E Series unit and follow steps to attempt to
retransmit.
•
Contact ZOLL Technical Service department if problem persists.
FAX NO CARRIER message is displayed.
Receiving phone line is not picking up or the sending modem is
experiencing problems. The E Series automatically attempts to
transmit until the transmission is successful or the E Series operator
presses the Abort softkey.
If retransmission fails:
•
Check the modem connection to the E Series unit.
•
Power cycle the cellular phone.
FAX NO DIAL TONE message is displayed.
•
Check the modem connection to the E Series unit.
•
Check the modem connection to the phone jack or cellular phone.
•
Power cycle cellular phone.
MODEM REQUIRED message is displayed.
No PCMCIA modem card is present. Insert PCMCIA modem card
and repeat transmission attempt.
Symptom
Recommended Action