Manual battery testing procedure, 0dqxdo %dwwhu\ 7hvwlqj 3urfhgxuh, Operation (quickcharge) – ZOLL M Series Defibrillator Rev G Base PowerCharger 4x4 User Manual
Page 13

9
Operation (QuickCharge)
To turn the ZOLL Base PowerCharger
4x4
on, plug the
power cord into live AC mains. All indicator lights will 
illuminate for approximately five (5) seconds and then 
go off. An audible tone will sound indicating self test 
passed and the POWER light will remain lit.
To charge a battery:
1. Check to see that the unit is connected to live AC 
mains. The POWER light should be lit.
2. Insert a battery into any compartment. Make sure the
battery is fully seated (a click can be heard).
3. The CHARGER ON indicator illuminates and remains
on during the charging cycle.
4. In four (4) hours or less for the PD4410/Smart/
Smart Ready Battery and eight (8) hours or less for 
all XL Batteries, the CHARGER ON light will go off 
and the BATT. READY light will illuminate indicating 
the battery has been charged to 100% of its present 
capacity and is ready for use.
If the battery or battery compartment is not capable of 
charging, the FAULT light will illuminate and the 
CHARGER ON light may stay on.
To identify the defect, press and hold the TEST 
button while the battery is still in the compartment. 
Illumination of certain indicator lights together 
represent a specific defect; see “Fault Codes and 
Definitions” on page 11 and take corrective action. 
Before disposing of the battery, test the battery 
compartment for defective operation (see “Battery 
Compartment Testing Procedure” on page 10) and 
perform a battery test. 
1RWH Interrupting a charge cycle in progress by
removing the battery will cause an audible 
warning of five (5) consecutive beeps to indicate 
that the battery is not fully charged and should not 
be used. In addition, all four indicator lights will 
illuminate for four (4) seconds and then go off. 
Repeated use of partially charged batteries will 
reduce battery capacity and operating life.
5. To remove the battery, insert a finger into the recess
at the top of the battery compartment, press against 
the battery to disengage the battery locking clip and 
lift the battery out. A compartment test is performed 
illuminating all indicator lights, briefly followed by the 
BATT. READY and TESTING indicator lights 
illuminating for two (2) seconds and then all indicator 
lights going off, except the POWER indicator light.
0DQXDO %DWWHU\ 7HVWLQJ 3URFHGXUH
ZOLL batteries are sealed lead-acid batteries. It is 
recommended that all batteries be tested for capacity at 
least every 90 days and that the test results and date be 
written on the battery label. 
All four battery compartments can test batteries. It is not 
necessary to insert a fully charged battery for testing; the 
ZOLL Base PowerCharger
4x4
will automatically charge
the battery first, test it, and then recharge the battery to 
full capacity. Testing a battery in one compartment will 
not affect any other charge or testing cycles in progress.
To test a battery:
1. Check to see that the ZOLL Base PowerCharger
4x4
is connected to live AC mains. The POWER indicator 
light should be lit.
2. Insert a battery into any compartment. Make sure the
battery is fully seated (a click can be heard).
3. Press the TEST button for the appropriate
compartment. The TESTING indicator light will
TEST
TESTING
FAULT
BATT. READY
CHARGER ON
TEST
TESTING
FAULT
BATT. READY
CHARGER ON
TEST
TESTING
FAULT
BATT. READY
CHARGER ON