REV 2 CT-7500 S2 USER’S MANUAL

4

Performing Diagnostics

Follow the steps below to perform diagnostics and confirm that the adapter is calibrated
properly:

a. Move the resistor transducer to one end of its travel.

b. From the “RUNNING” screen, press either the [▲] or [▼] key to access the Main Menu.

The following screen will be displayed:

The DIAGNOSTIC option should be highlighted. Press the [ENTER] key.

c. The following screen will be displayed:

Move the resistor transducer and observe the distance values on the screen:

Move the resistor transducer to the other end of its travel and observe the screen:

The value should be the distance between the two endpoints of the transducer. If this
value is not correct, you should calibrate the adapter.

ENCODER TRAVEL:

5.00 IN 12.7 CM

“ENTER” WHEN DONE

ENCODER TRAVEL:

2.23 IN 5.66 CM

“ENTER” WHEN DONE

ENCODER TRAVEL:

0.00 IN 0.00 CM

“ENTER” WHEN DONE

CALIBRATE ENCODER

ADJUST CONTRAST