CT-7500 S2 USER’S MANUAL

REV 2

1

Calibrating the Resistor Transducer Adapter

The Vanguard Resistor Transducer Adapter must be calibrated each time a different resistor
transducer is connected to the adapter. Follow the steps below to calibrate the adapter:

a. Connect the resistor transducer to the adapter.

b. Turn on the circuit breaker timer. The following “RUNNING” screen will be displayed on

the resistor transducer adapter’s LCD screen:

Press either the [▲] or [▼] key to access the Main Menu.

c. The Main Menu will be displayed:

Press the [▲] key to select the CALIBRATE ENCODER option:

Press the [ENTER] key.

d. The following screen will be displayed:

Move the resistor transducer to one end of its travel and then press the [ENTER] key.

e. The following screen will be displayed:

Move the resistor transducer to the opposite end of its travel and then press the
[ENTER] key.

SET ENCODER TO
POSITION 2

“ENTER” WHEN DONE

SET ENCODER TO
POSITION 1

“ENTER” WHEN DONE

ADJUST CONTRAST

DIAGNOSTIC

CALIBRATE ENCODER

ADJUST CONTRAST

RUNNING


Res Enc Rev: 1.00