Calibrating the Resistor Transducer Adapter
The Vanguard Resistor Transducer Adapter must be calibrated each time a different resistor
transducer is connected to the adapter. Follow the steps below to calibrate the adapter:
a. Connect the resistor transducer to the adapter.
b. Turn on the circuit breaker timer. The following “RUNNING” screen will be displayed on
the resistor transducer adapter’s LCD screen:
Press either the [▲] or [▼] key to access the Main Menu.
c. The Main Menu will be displayed:
Press the [▲] key to select the CALIBRATE ENCODER option:
Press the [ENTER] key.
d. The following screen will be displayed:
Move the resistor transducer to one end of its travel and then press the [ENTER] key.
e. The following screen will be displayed:
Move the resistor transducer to the opposite end of its travel and then press the
[ENTER] key.
SET ENCODER TO
POSITION 2
“ENTER” WHEN DONE
SET ENCODER TO
POSITION 1
“ENTER” WHEN DONE