beautypg.com

Manual deskewing procedures, Cable deskewing using the fast edge output – Teledyne LeCroy QPHY-HDMI User Manual

Page 31

background image

QPHY-HDMI Operator’s Manual

Manual Deskewing Procedures

This section is only applicable to the oscilloscope and the cables connecting to the oscilloscope
channels.

Cable Deskewing Using the Fast Edge Output

The following procedure demonstrates how to manually deskew two oscilloscope channels and
cables using the fast edge output, with no need for any T connector or adapters.

Note: Fast Edge output is available only on some models. If your oscilloscope does not have
Fast Edge output, see

Cable Deskewing Using the DUT Signal

.

This can be done once the temperature of the oscilloscope is stable. The oscilloscope must be
warmed up for at least a half-hour before proceeding. This procedure should be run again if the
temperature of the oscilloscope changes by more than a few degrees.

For the purpose of this procedure, the two channels being deskewed are referred to as Channel
X and Channel Y. The reference channel is Channel X and the channel being deskewed is
Channel Y.

1. Begin by recalling the Default Oscilloscope Setup.

2. Configure the oscilloscope as follows:

• Timebase

i. Fixed Sample Rate

ii. Set the Sample Rate to 40 GS/s

iii. Set the Time/Division to 1 ns/div

• Channels

i. Turn on Channel X and Channel Y.

ii. Set V/div for Channel X and Channel Y to 100mV/div.

iii. Set the Averaging of Channel X and Channel Y to 500 sweeps.

iv. Set the Interpolation of Channel X and Channel Y to Sinx/x.

922540 Rev A

27