Use with skc rotameters, Use with other sampling heads, Use with other types of calibrator – SKC Limited IOM Sampler User Manual
Page 2
Page 2
007-09-008 Issue D October 2009
SKC Limited 11 Sunrise Park, Higher Shaftesbury Road, Blandford Forum, Dorset DT11 8ST United Kingdom
T: +44 (0) 1258 480188 F: +44 (0) 1258 480184 E: [email protected] W: www.skcltd.com
Use with SKC rotameters
The Calidaptor is screwed into the top
(outlet port) of the rotameter, tightening
enough to just compress the ‘O’ ring
sealing it to the rotameter.
Pass the sampling head through the
hinged bracket, and rest its inlet against
the foam ring. Clamp the sampling head
in place with the plastic clamping screw
until the foam ring is compressed by
approximately 1mm.
Connect the sample head outlet hosetail
to the pump inlet with a length of tubing.
Use with other sampling heads
With the Calidaptor fitted to the top of
an SKC rotameter, screw the supplied
hosetail fitting into the threaded hole
inside the Calidaptor bowl. Connect
from the hosetail to the inlet of the
sampling head with a length of flexible
tubing.
Note: If subsequently calibrating an
I.O.M. (or similar) sampler, especially
if used with a foam insert, ensure that
the hosetail is removed from inside the
Calidaptor bowl.
Use with other types of calibrator
The Calidaptor can be used as
described above, with any other type
of flow calibrator, by screwing the
supplied hosetail into the threaded hole
in the inlet of the Calidaptor. Connect
from the hosetail to the flow calibrator
with a length of flexible tubing.