Troubleshooting – SKC 223-Series PowerFlex - Single User Manual
Page 3
3
5. The LED at the port will light a steady red to indicate that a charge sequence
has been initiated.
If the LED does not light to red in < 60 seconds, see Troubleshooting,
Step 1.
6. Once the batt ery pack is fully charged, the LED at the port will change to a
steady green to indicate that PowerFlex has switched to a trickle charge to keep
the batt ery at full charge.
7. The batt ery can be removed at any time aft er the LED turns green. Detach the
batt ery from the cable and remove the cable from the charger.
Important Note: Pumps may remain connected to the charger indefi nitely
(without causing damage to the pump or charger).
Important Note: Remove pump-specifi c cable from charger when not in
use to avoid shorting.
Troubleshooting
If the LED does not light when a batt ery pack is plugged into the port or the LED
does not turn off aft er a batt ery pack is unplugged, follow the steps in the table.
Step
Action
No result, Action
Yes result, Action
1
Keep the unit powered on. If the battery is
deeply discharged, the LED may take up to
30 minutes to light or turn off.
Go to Step 2
Proceed with operation
2
Move unit to a different wall outlet
Go to Step 3
Proceed with operation
3
Unplug charger from its power source for
20 seconds. Plug it back into power source.
Go to Step 4
Proceed with operation
4
Replace pump-specifi c cable
Go to Step 5
Proceed with operation
5
Try a different battery pack
Call SKC
Proceed with operation —
previous battery pack
may be damaged,
call SKC.