E Using an Intalogix™ Technology interface (QMB 15291):
QMB
Intalogix™
Terminal
Wire Color
Function
Technology Inst
1
Green
(-) Excitation
1
2
Red
(+)Excitation
2
3
Black
Ground
3
4
White D out
4
5
Brown
D In
5
6 Blue EOC
6
7
Orange
SCLK
7
8
Yellow CS
8
9 Violet
Temperature
9
10 Gray/Shield
Chassis
10
7. Adjust the analog interface indicator to the platform. First, ensure all corners are
within 1 division of each other at 25% of rated capacity. Follow the appropriate
indicator service manual to ensure a good calibration.
To do this:
A. Perform a coarse platform calibration which is close to the actual weight.
B. Identify the platform's corner numbers.
50637 / SJ4404
6
08/02 Issue 2
Note:
Calibrate the Intalogix™ platform/ indicator. Follow the appropriate indicator service manual to
ensure a good calibration.
Top View
50224-1
2
3
4
1