E Using an Intalogix™ Technology interface (QMB 15291):

QMB

Intalogix™

Terminal

Wire Color

Function

Technology Inst

1

Green

(-) Excitation

1

2

Red

(+)Excitation

2

3

Black

Ground

3

4

White D out

4

5

Brown

D In

5

6 Blue EOC

6

7

Orange

SCLK

7

8

Yellow CS

8

9 Violet

Temperature

9

10 Gray/Shield

Chassis

10

7. Adjust the analog interface indicator to the platform. First, ensure all corners are

within 1 division of each other at 25% of rated capacity. Follow the appropriate
indicator service manual to ensure a good calibration.

To do this:
A.
Perform a coarse platform calibration which is close to the actual weight.

B. Identify the platform's corner numbers.

50637 / SJ4404

6

08/02 Issue 2

Note:
Calibrate the Intalogix™ platform/ indicator. Follow the appropriate indicator service manual to
ensure a good calibration.

Top View

50224-1

2

3

4

1