Welch Vacuum 8925 Direct Drive Rotary Vane Pump User Manual
Page 22
22
11. REPLACE THE LIP SEAL HOUSING
a) It is recommended that you use 3 lockwashers to secure the LIP SEAL HOUSING. Uniform
tightness is important. Repeatedly finger-tight the 3 screws in sequence.
Then lastly, turn them very tight.
b) Replace the Woodruff Key in its slot. Use your small finger or a thin tool to hold it in place as
you replace the motor coupling housing. Be sure the key stays in position. If it should fall even
a little distance, use a screwdriver to push it back up in the key slot.
Then tighten the set key screw—very tight.
12.
REPLACE OTHER COMPONENTS then reassemble the motor and other components.