Welch Vacuum 8925 Direct Drive Rotary Vane Pump User Manual
Page 17
17
Section 5: OPERATION
5.01 Starting Procedure
Before using the pump for the first time, it is a good idea to spend a few minutes inspecting the pump
and its electrical and vacuum connections. Review Section 1: Installation as required.
Check the AC power outlet to be sure that it is the same voltage and phase as the pump motor.
Connect the power cord to the power outlet and recheck the oil level.
Close off the pump intake and the gas ballast valve, and run the pump at blankoff for a few minutes.
The gurgling noise should go away after a few minutes of running; it is caused by the high volume of
air that flows through the pump when the pump is first turned on. If the gurgling noise does not stop,
check the oil level to see if it is low, and check the pump intake fitting to be sure that it is tight. Once
proper pump operation has been verified, the pump intake can be opened to the vacuum system.
After running the pump for a few minutes, check the oil level again. If the level is too high or too low,
stop the pump and add or remove oil as needed. Stop the pump and vent it to the atmosphere before
adjusting the pump fluid level.
Before starting the pump when connection to the vacuum system, check all vacuum connections.
5.02 High Pressure Operation
The Vacuum Pumps are designed to be most efficient when operated at or near their ultimate blankoff
pressure. When operated at elevated pressures for long periods of time, the pump will run hotter.
Additional cooling may be required for both the pump and the motor. At elevated ambient temperatures
under these conditions, the thermal protection switch may cut out. The oil may tend to thin out at these
temperatures, losing its ability to lubricate effectively.
During operation at these pressures, oil mist may be generated at the exhaust port. The mist has the
appearance of a small cloud of smoke. Use of an Exhaust Filter with coalescing element is required. An
exhaust filter with an oil drain back feature is highly recommended. See Section 9: Accessories.
5.03 Shutdown Procedures
A few simple precautions are necessary before performing a pump shutdown. If a gauge is connected to
the system, first isolate the gauge, then turn off the power to the pump and open the system to the
atmosphere.
If the pump is disconnected from the system for any length of time, cover the pump intake with a rubber stop-
per or other suitable cover to protect the pump against contamination. An intake screen is furnished to prevent
loose particles from entering.
If the exhaust port is open, that should also be covered. If the pump oil is contaminated and the pump is go-
ing to be stored for a prolonged period, the oil should be changed before the pump is stored. Even if a pump
is stored for a long period with oil initially in good condition, check the oil when the pump is restarted, and
change the oil if necessary.
NOTE
The intake isolation valve will automatically close when power
to the pump is turned off. This will maintain vacuum in the system
(if the pump remains connected to the rest of the system).
NOTE
If operating this pump for prolonged periods of time at or
above 10 Torr,contact the Welch Vacuum Technical Service
Department for further information and precautions.