Model GFC7000T Carbon Dioxide Analyzer

Teledyne Analytical Instruments

1

1. OVERVIEW

The Model GFC 7000T is a close derivative of the Model GFC 7000TA CO

2

Analyzer. The

main differences are as follows.

The optical bench is longer (14 m), making the instrument more sensitive at low CO

2

levels. The instrument has a user-selectable full scale range of 0-100 PPB to 0-100 PPM
of CO

2

.

The rear panel is different. There is an extra pneumatic port for the exhaust of the purge
gas to the GFC wheel. Also, there are relay alarm outputs available on the rear panel.

Refer to the main manual (i.e. GFC 7000TA) that is shipped with your instrument for
details of the instrument’s components and operation. Any differences between this
instrument and what is described in the GFC 7000TA, will be described in this Addendum
to the manual.

2. SPECIFICATIONS AND AGENCY APPROVALS

Table 2-1 presents the specification parameters and values, as well as agency approvals.

Table 2-1. GFC 7000T Specifications and Agency Approvals

PARAMETER SPECIFICATION

Ranges

User selectable to any full scale range from 0-2 ppm to 0-2,000 ppm

Measurement Units

ppb, ppm, µg/m

3

, mg/m

3

(user selectable)

Zero Noise

< 0.1 ppm RMS

Span Noise

< 1% of reading RMS

Lower Detectable Limit

< 0.2 ppm

Zero Drift (24 hours)

< 0.25 ppm

Span Drift (24 hours)

< 0.5% of reading

Linearity

1% of full scale

Precision 0.5%

reading

Lag Time

10 sec

Rise/Fall Time

<60 sec to 95%

Sample Flow Rate

800 cm

3

/min. ± 10%

Temperature Range

5 - 40

C operating

Humidity Range

0-95% RH, Non-Condensing

Temp Coefficient

< 0.05 % of reading per

C

Voltage Coefficient

< 0.05 % of reading per V

Dimensions (HxWxD)

7" x 17" x 23.5" (178 mm x 432 mm x 597 mm)

Weight

40 lb (18.1 kg)

AC Power

100 – 120V 50/60 Hz,