7. System Maintenance

Gas Processing:

Attention:

Empty the condensate reservoir completely after each measuring

operation. Water residues within the measuring instrument will destroy the
pumps and sensors!
Damage of the filter and / or improperly fitted filter will greatly decrease o

eliminate the filter function andwilleventually destroy pumps and sensors.

Plug-type elements
and flanges:

Storage:

See drawing on page 44.

Check themicrofilter for contaminations and replace as necessary.

If the pump capacity is reduced, exchange the diaphragm filter.

Make sure that threaded parts are straight when placed on and tighten them
moderately. Ensure sufficient sealing bymeansofO-rings.

Remove any gas residues. Grease with Vaseline.

Store in a cool and dry environment at a temperature of approx. 60 °F (20 °C).