beautypg.com

Teledyne 9060L - Zirconium oxide flue gas oxygen analyzer User Manual

Page 25

background image

Percent Oxygen Analyzer

Installation

Teledyne

Analytical

Instruments

25

probe or install a longer probe. This is normally caused by stratification
of the flue gas.

Figure 3-1: Oxygen Probe Mounting

3.4 Electrical Connections

Figure 3-1 shows the Model 9060L rear panel. For detailed

pinouts, see the wiring/interconnection drawings in the Drawings section
at the rear of this manual.

3.4.1 Primary Input Power

The power cord receptacle, fuse block and Power switch are

located in the same assembly. A 6-foot, standard AC power cord is
supplied with the control unit. Insert the female plug end of the power
cord into the power cord receptacle.

The power supply allows direct connection to any 110 VAC,

50/60Hz power source. 220VAC is an optional configuration settable
through an internal switch in the control unit. The fuse block, to the right
of the power cord receptacle, accepts two 5x20mm IEC type T, 2 A,
time-lag (T) fuse. (See Fuse Replacement in chapter 5, Maintenance.)

The Power switch is located on the right-hand end of the power

source input receptacle assembly.