Mini-GASS User Manual
6. If system is being installed with automatic filter blow back option, install blow back
system on right flange bolt (refer to Figure 16).
7. Attach pre-bent SS tubing to Mini-GASS and to blowback assembly.
8. Lift Mini-GASS system into place so union fitting on back of system can be attached to
probe pipe. This is most easily accomplished with help of an assistant.
9. While holding system in place, tighten union hand tight.
10. Connect free end of turnbuckle to anchor point at middle back of system (refer to Figure
17).
11. Attach with stainless clevis pin.
12. Snap keeper wire into place.
13. Adjust turnbuckle by turning it out to increase length until union fitting is straight.
Turnbuckle is used only to relieve torsional load from union fitting.
14. Tighten union fitting with a wrench.
15. Install insulation between flange and back of Mini-GASS.