6. TROUBLESHOOTING GUIDE

PROBLEM CAUSE

SOLUTION

No display

a) Main power not

switched on.

b) AC Adapter

socket not
inserted
properly.

a) Switch on the power supply.

b) Re-insert AC Adapter & press

.

a) Insufficient

reference
electrolyte in
electrode.

a) Fill electrode with reference electrolyte.

b) Broken

electrode.

b) Replace electrode.

c) External ‘noises’

or induction
(e.g. electrical
‘noise’ caused
by a nearby
motor).

c) Remove or switch off interfering device.

Unstable
reading

d) Dirty electrode.

d) Clean & rehydrate the electrode.

Slow
response

a) Dirty electrode.

a) Clean & rehydrate the electrode.

Meter not
responding
to key press

a) HOLD mode

in operation.

b) Bad keypad.

a) Press

to deactivate. See SETUP

to disable Auto Hold feature if enabled.

b) Contact Technical Service.