PROBLEM CAUSE
SOLUTION
CAL
% calibration value is not
acceptable.
Electrode is disconnected
or failing.
Check electrode connections.
Check that correct DO unit is
selected & repeat cal procedure.
Service electrode.
Reset meter.
Replace electrode.
CAL
ppm or mg/L calibration is
attempted when the factory
calibrated absolute value is
below 2.00
Use concentration calibration
solution higher than 2.00.
Check electrode connections.
Check that correct DO unit is
selected & repeat cal procedure.
Service electrode.
Reset meter.
Replace electrode.
Slow
response or
unstable
readings
External noise pickup or
induction caused by
nearby electric motor
Insufficient electrolyte in
probe
Air bubbles near sensor
Dirty or damaged probe
Probe not deep enough in
sample
Broken probe
Remove or switch off interfering
device
Fill probe with electrolyte & / or
replace pre-assembled cap
Stir or tap probe to remove
bubbles, reduce stirring
Clean the probe and re-calibrate
Submerse sensor cap into sample
Replace probe or membrane
25
- B4000-16 (28 pages)
- B4000-M (23 pages)
- Foxx Life Sciences (2 pages)
- CON 700 (24 pages)
- pH-CON 700 (32 pages)
- 2700 Series Benchtop Meters (60 pages)
- Oakton Testr 35 (6 pages)
- 4066, 4067, 4068, 4069, 4161, 4162, 4173, 4174, 4176, 4177 Control Company (4 pages)
- 4065, 4175, 4270, 4271, 4274, 4565, 4570, 4574 Control Company (4 pages)
- 4170 Control Company (4 pages)
- WD-35653-15 (3 pages)
- WD-35653-13 (1 page)
- pH-CON 300 (26 pages)
- pH-CON 10 (29 pages)
- CON 6+ (40 pages)
- 4070 Control Company (1 page)
- WD-00653-00 Zero Oxygen (1 page)
- pH-Dissolved Oxygen Meter (35 pages)
- DO 6+ (48 pages)
- Foxx Autofil Bottle Top Filtration (2 pages)
- Foxx Autofil Centrifuge Funnel and Stand (2 pages)
- Foxx Autofil Laboratory Storage Bottles (2 pages)
- Foxx Autofil Pedestal and Cradle Ring (4 pages)
- RITEFLOW FLOWMETERS AND KITS (16 pages)
- Roto-Flo Flow Indicators 19937 Series (1 page)
- Leveling Base Kit (each) H40404-1000 (1 page)
- Taylor M-3000 TTi3000 Colorimeter Series (28 pages)
- Bel-Art Magnetic Stirring Bars Selection & Use Guide (4 pages)
- VISPOL-Series Polarimeters (2 pages)
- UniPol L-Series Polarimeters (4 pages)
- iPR-Series (8 pages)
- Atago AP-300 Series Polarimeter (4 pages)
- Foxx Heavy Duty Vacuum Bottle (2 pages)
- pH 700 Ion 700 (18 pages)
- pH Tutor Benchtop Meter (2 pages)
- pH 4.01 Buffer (1 page)
- pH 5+ (28 pages)
- pH 11 (76 pages)
- pH 300 (24 pages)
- 4080 TRACEABLE HUMIDITY_TEMPERATURE DEW POINT METER (2 pages)
- 4088 TRACEABLE HYGROMETER_THERMOMETER (1 page)
- 3651 TRACEABLE ANEMOMETER_HYGROMETER_THERMOMETER_DEW POINT PEN (2 pages)
- 3652 TRACEABLE ANEMOMETER_HYGROMETER_THERMOMETER_DEW POINT BAROMETER PEN (3 pages)
- Sibata 6168 Filter Holders (1 page)