Nova-Tech pH-CON 10 User Manual
Page 10
![background image](/manuals/340075/10/background.png)
Instruction Manual
PC 10
- 6 -
4 . C A L I B R AT I O N
4.1.
Important Information on Meter Calibration
When you recalibrate your meter, old pH and conductivity calibration data are replaced on a
point by point basis. For example, if you previously calibrated your meter at pH 4.0, 7.0, and
10.0, and you recalibrate at pH 7.0, the meter retains the old calibration data at pH 4.0 and pH
10.0.
To completely recalibrate your meter, or when you use a replacement probe, it is best to set
the meter to its factory defaults and recalibrate the meter at all points.
To reset the meter to its factory defaults:
1. While in Measurement mode, press
CAL/MEAS and hold for 3 seconds.
2.
The meter will prompt RST in the lower display
and CAL in the upper display. See Figure 6.
3.
Press ENTER to reset your meter to its factory
defaults. If you do not want to erase existing
calibration data, press CAL/MEAS to escape
this mode.
For information on how to calibrate your meter:
See Section 4.3 on page 8 for Temperature calibration.
See Section 4.4 on pages 10 for pH calibration.
See Section 4.5 on pages 12 for conductivity calibration.
Figure 6: Reset to factory defaults