Nova-Tech WBW Series User Manual
Page 18
© Adam Equipment Company 2009
16
Method 2
Alternatively, you may calibrate entering the group 2 Parameters (see
Section 13.2).
• Press [Tare] when “S6 CAL” is displayed.
• Display will show “UnLoAd”.
• Remove all weights and press [Tare] when the scale is stable.
• Display will show “Ld xx”. Place the suggested calibration mass on the
pan.
• Press the [Tare] key when the stable indicator is on.
• The scale will calibrate to the mass and then return to the parameter
“S6 CAL”
• Press the [Zero] key to return to weighing.
• Remove the calibration weight as soon as calibration is complete.
NOTE: If an error message “FAIL H” or “FAIL L” is shown during calibration, re-
check the calibration and repeat, if necessary. If the error cannot be
corrected contact your dealer or Adam Equipment for advice.