Multichannel Systems TC01-TC02 Manual User Manual
Page 16

Temperature Controller TC01/02 Manual
12
Diagnosis 2: Controller settings
This diagnose screen view is used for reviewing and checking the user settings.
Setpoint Temp
Setpoint temperature
P Gain
Proportional gain
I Gain
Integrator gain
Max Power
Maximum output power
Diagnosis 3: Controller output
This diagnose screen view is used for checking the operation of the internal controller.
Power Set
Output power set by the controller.
Power Out
Actual output power (product of Current Out and Supply Voltage)
Duty Cycle
PWM Duty cycle (internal value)
Current Out
Current through the primary coil of the isolating transformer
Supply Voltage
Supply voltage (from power supply)
Diagnosis 4: Heating element
This diagnose screen view is used for checking the connected heating element.
On/off
Current channel state
HE Voltage
Output voltage applied to heating element
HE Current
Output current applied to heating element
HE Resistance
Heating element resistance (voltage-current ratio)
HE Power
Output power delivered to heating element (voltage-current
product), should be 80 - 90 % of Power Out, depending on
heating element resistance)