or the high pressure line is below atmospheric pressure

they should be checked for leaks or plugs. Check for

proper zero setting.

Low Reading — Specific gravity of indicating fluid may

be too great: check scale legend and provide correct

indicating fluid. Check for leaks or plugs in high pressure

line, if it is above atmospheric pressure, and or in low

pressure line, if it is below atmospheric pressure. If meter

has a valve by-pass line, this valve must be leak free.

Check vent when used. Check for proper zero setting.

10. Maintenance

Normal maintenance necessary with Meriam manometers

is the occasional cleaning of the tubes. Some manometer

indicating fluids become oxidized with use. Others, over

long periods of time, react with gases and fluids with

which they are in contact, leaving a deposit on the surface

of the fluid, and, in turn, on the tube.

Prior to cleaning the glass tube, a manometerʼs indicating

fluid must be drained using the steam flow drain valve on

well type manometers or by carefully tipping a U-tube

manometer on its side. As the fluid in the manometer

may be classified as a SARA Title III Section 313 Toxic

Chemical and may be harmful to your health if handled

improperly, please consult the Material Safety Data Sheet

(MSDS) sheet supplied with this product for detailed

health and safety information and proper handling

information. Dispose of used indicating fluids and used

cleaning materials according to local, state, and federal

regulations.

The method of cleaning the various models is similar,

but, there are some differences. It is generally more

convenient to clean from the upper end of the manometer.

On installations where accessories or piping is connected

to the upper end of the manometer and it is inconvenient

to remove equipment, the tube can often be cleaned

through the drain plug connection in the lower channel

end block. It is recommended that wherever possible the

instrument be removed to a convenient table where it can

be laid horizontally to afford greatest care in performing

the various operations.

Meriam Cleaning Fluid

A clean manometer is essential for readability

and accuracy. Meriam No. 915E cleaning fluid is

recommended for use in all Meriam manometers. It is

available in gallon containers only.

11. Parts

For replacement and/or spare parts request parts

drawing specifying the following information.

Model number

Serial number

Range

12. Accessories

To broaden the application and simplify operation

of your instrument, Meriam provides the following

accessories:

Return Wells— for over-range protection

Fluid Check Valves — for over-range protection

Line Traps — to keep instruments clean by trapping

dirt from the process.

Moisture & Sediment Traps — to keep instruments

clean by trapping dirt from the process

Pressure Vacuum Variator.— static pressure/vacuum

pressure

Valves & Piping Manifolds — to simplify different

pressure measurements

Sight Feed Bubblers — for metering purge gasses

Table Mount Stands

Cleanout Bore Brushes

Seal Pots — for isolating process fluid from the

instrumentation

Orifice Plates & Flanges

Meriam Instrumentation has pioneered the

development and application of manometry since

1911. The highest standards of accuracy and

practical, high quality design have made the words

MERIAM and MANOMETERS synonymous to

the engineering and research world. We welcome

your inquiries pertaining to manometry, and will be

pleased to help in any way we can.