When you decide to have your LFE cleaned, please be aware of the various capabilities of calibrating

LFE’s at Meriam. Every calibration includes: 6-8 calibration points [differential pressure 1”-8”H2O], a data
sheet with raw and reduced data, calibration curve, and instruction manual. The procedure numbers are
noted in (parentheses).

Calibrations

NIST Traceable Certificate Accuracy

(% Reading)
( )

Clean per (A33555) and calibrate per (A35822) Working-Master

X

±0.72% R

( )

Clean per (A33555) and calibrate per (A35822) Master-Master

X

±0.54% R

( )

As Received calibration (A35822/A34777), clean, and calibrate [WM] X

±0.72% R

( )

As Received calibration (A35822/A34777), clean, and calibrate [MM] X

±0.54% R

( )

Accredited calibration ISO/IEC 17025 (contact Sales for details)

X

±0.64% R

( )

Nuclear or safety application (A33544)

X

±0.72% R

( )

Subcontracted per (A35352)

X

±0.50% R


Options
( )

2 additional calibration points beyond full scale- up to 12”H2O

( )

Additional calibration points beyond full scale at the following settings _____________________.

( )

Additional calibration points below 1”H2O at the following settings _______________________.

( )

Oxygen cleaning per (A50558)

( )

Calibrating the LFE with differential pressure transmitter

( )

Hydrostatic leak testing per (A33559)

( )

Pneumatic pressure test per (A70763)


The standard reference unit for flow rate is (cfm) cubic feet per minute. The additional units available are:
( )

Liters

( )

Cubic Centimeters

( )

Meters

( )

Pounds (include flowing temperature and pressure)

( )

Kilograms (include flowing temperature and pressure)

( )

Other _____________________


Time constants:
( )

Second

( )

Minute

( )

Hour


The standard reference for differential pressure is inches of water at 4°C. The additional units available are:
( )

Millimeters of water @ 4°C

( )

Centimeters of water @ 4°C

( )

Pascals

( )

Kilopascals

( )

Other _____________________


A second data sheet and calibration curve using the listed units will be included with the calibration